Infant Ventilator AM-Y200

Category:

Description

 

Introduction

AM-Y200 ventilator is a new-type of pneumatic-driven and electronic-controlled ventilator which is designed to meet the special needs of clinical respiratory management and respiratory therapy for infants and children with respiratory disorders.

With the feature of innovative functions, convenient operation, less gas consumption, compact but reasonable structure, and outstanding safety and reliability, it is an indispensable instrument in emergency and ICU rooms, and an essential tool to rescue the life of infants and children.

-AR-Y200 ventilator can work with some innovative ventilation modes, e.g., constant flow

mode, pressure limit mode and time cycle ventilation mode, and also supports PEEP and IMV functions.

Technical Specifications

PARAMETER SPECIFICATION
Working Mode (Ventilation Mode) Controlled Mechanical Ventilation (CMV),

Intermittent Mandatory Ventilation (IMV),

Continuous Positive Airway Pressure (CPAP),

Manual Ventilation (Manual).

Breath Rate 1~120BPM
Inspiratory to Expiratory Ratio 1:1~1:3
Special Order I:E Range 10:1~1:9.9
Inspiratory Flow 3 L/min~20 L/min
Tidal Volume Minimum 10 mL~300 mL
Minute Volume Max. Ventilation volume greater than 5 L/min
Oxygen Concentration of Output Gas 21%~100%
Ventilator System Compliance Not greater than 3×10-2 mL/Pa
Positive End-Expiratory Pressure (PEEP) 0kPa~2.0kPa
Inhalation Safety Valve Safety valve shall be opened when the airway pressure is within -300 Pa ~ -800 Pa
Inspiratory Pressure 1kPa~9kPa
Inspiratory Time 0.2s~3.0s
Synchronous Trigger Pressure Trigger -1.0~0kPa (Special Order)
Flow Trigger 0~10L (Special Order)
Maximum Safe Pressure Not greater than 10kPa
Continuous Positive Airway Pressure (CPAP) 0kPa~2.0kPa
Monitoring Tidal Volume Real-time monitoring within the adjustment range of tidal volume: <100 mL: ±10 mL; ≥100 mL: ±15 %;
Monitoring Oxygen Concentration 21%~100%
Power Failure Alarm Audible alarm sounds immediately in case of power outage and lasts at least for 120s
Gas Source Failure Alarm Audible and visual alarm sounds within 6s after gas source interruption occurs and lasts at least for 120s
Airway Pressure Lower Limit Alarm Range PEEP +0.2kPa
Airway Pressure Upper Limit Alarm Range Inspiratory pressure +0.4 kPa