Horizontal Cylindrical Pressure Steam Sterilizer AM-WS-YDC

Characteristics

  1. The steam can’t go through to chamber when the door is not close, and the pilot off inform. You can process dry operate.
  2. When the pressure inner over 0.027MPa, the door be locked and can’t be open.
  3. If you not choose the sterile time and start the machine, buzzer sound.
  4. If you mistake open the door in the course of start, the jacket value auto off and open the chamber value.
  5. Lack water the pilot on, heating and sterile pilot off, the buzzer sound and auto release steam.
  6. Overload and leakage protection, when leakage or over current occur, it auto off the power.
  7. Temperature controlled not work and the inner pressure over 0.165MPa, the safety value auto start opens.

Description

Technical Data

Model technical data WS-280YDC WS-200YDC WS-150YDC
Sterilizing chamber volume 280L(φ600×1000mm) 200L(φ515×1000mm) 150L(φ440×1000mm)
working pressure 0.22 MPa
working temperature 134℃
Adjustment of temperature 40℃ – 134℃
Time of sterilization 0-60 min
Time of drying 0-60 min
Heat average ≤± 2℃
Power 12KW/380V 50Hz 9KW/380V 50Hz 9KW/380V 50Hz
Dimension 1400×770×1780 1400×670×1650 1400×600×1300
Package 1680×920×2100 1560×820×1850 1550×750×1850
GW/NW 465/365 KG 350/260 KG 320/240 KG

Reviews

There are no reviews yet.

Be the first to review “Horizontal Cylindrical Pressure Steam Sterilizer AM-WS-YDC”

Your email address will not be published. Required fields are marked *