Electrosurgical Unit AM-ESU-L2000 LEEP
Description
Specifications
-Microcomputer control
-ESU-L2000 (LEEP) generator adopts LEEP which specializing in gynecological intrauterine event, increasing the accuracy, controllability, security of the surgery.
-Dealing with some kinds of pathological changes in obstetrics and gynecology, for example, cutting, electrocoagulation, cervical erosion, and some kinds of surgery that fit in peritoneoscope.
-Different electrodes have different effects in coagulation.
-An optional smoke evacuation system is available.
-Particular illuminating ESU pencil can increase the clarity of wound healing.
Technical Specifications
PARAMETER | SPECIFICATION | |
Power | 220V±22V, 50Hz±1Hz | |
Operating Frequency | 360~460kHz | |
Power Rating | 880VA±10% | |
Five Working Modes | Monopolar Cut | a) Cut1 (pure cut): 1W~200W (Load 500Ω)
b) Cut2 (Blend1): 1W~150W (Load 500Ω) c) Cut3 (Blend2): 1W~100W (Load 500Ω) |
Monopolar Coag | d) Coagulation1 (Spray Coag): 1W~120W (Load 500Ω)
e) Coagulation2(Forced Coag): 1W~80W ( Load 500Ω) |
|
Power Consumption | ≤1100VA | |
SURGICAL SAMOKE EVACUATION | ||
Blowing Rate | 100 l/min @ 6mm tube | |
Noise | <60dB(A) | |
Start Control | Manus/Foot/Synchronize | |
Suction Adjust | 10%~100% | |
Stop Delay Time | 0~60s | |
AC Power | AC90~250V, 50Hz | |
Power Consumption | < 350W | |
Filter Characteristics | 99.999% @ 0.1-0.2μm | |
Active Carbon | 97% @ 1 year |
Packing List
Mainframe 1set
Electrosurgical pencil 3pieces
Illuminating electrosurgical pencil 2pieces
Electrosurgical pad 10 pieces
Electrosurgical pad cable 1piece
Foot switch 1set
Electrodes 1case(10pieces)
Tip clean sponge 1piece
Fuse 3pieces
Power cord 1piece
Operating manual 1book