6 Channel ECG Machine With Color Screen AM-ECG-C06G
Features
- Compact in size, easy and direct operation, portable
- Accurate pacing pulse identification and automatic analysis function.
- Four working modes: Manual / Auto / Physical Examination / Storage
- 112mm, 6 channel format recording
- Freeze function: can replay 12 seconds cardiograph front and back
- Fast ECG test under the [Physical Examination] ,saving power and paper
- Store and replay cardiograph without printing parameters and diagnosis under the Storage Mode.
- 7”800×480 graphic 7 inch touch screen, fast input patient information – Adapt to 110-230V, 50/60Hz AC power supply..
- Built in rechargeable Li-ion battery.
- 12 lead waveforms view with 250 patient information storage (Extendable)
- Built-in USB,RS232 interface , EXGNET software optional
- USB interface for net transmission, USB flash disk and USB printer.
Description
Technical Specifications
PARAMETER | SPECIFICATION |
Lead | Standard 12 leads |
Lead acquisition | 12bit/1000Hz(synchronously 12 leads) |
Work mode | MAN, AUTO, ANALYSIS Modes |
Filter | AC Filter : 50Hz / 60Hz
EMG Filter : 25Hz / 45Hz Anti-Drift Filter : 0.15Hz (Adaptive) |
CMMR | ≥100dB (with AC FILTER) |
Input circuit | Floating; Protection circuit against
Defibrillator effect |
Input Impedance | ≥50MΩ |
Input circuit current | ≤0.05μA |
Patient current leakage | <10μA |
Calibrating voltage | 1mV±2% |
Voltage tolerance | ≥±500mV |
Time constant | ≥3.2s |
Frequency response | 0.05~160Hz(-3db) |
Noise level | <15μVp-p |
Inter-channel
Interference |
≤0.5mm |
Sensitivity | 5,10,20, 40, Auto (mm/mV) |
Recording mode | 3ch, 3ch+, 3ch+++, 6ch
Default Man. Mode is 3ch format Default Auto. Mode is 6ch+ format |
Paper speed | 6.25、12.5、25、50mm/s(±3%) |
Paper | 112mm *20m Roll paper |
Communication | RS232/ USB port |
Power Supply | AC 110-230V( ±10%),50/60Hz ( ±1Hz),60VA
DC Built-in rechargeable battery,14.4V(2000mmAh) |
Dimension/Weight | 300mm×230mm×72mm/2.8Kg |
Reviews
There are no reviews yet.